Ion-selective electrodes (ISE, ISFET)FR Électrodes sélectives d'ions (ISE, ISFET) (Nitratomètre, ionomètre de poche)TR İyon seçici elektrotlar (ISE, ISFET) (Nitrat ölçer, potasyum ölçer, cep iyon metresi)
Also called: Nitrate meter, potassium meter, pocket ion meter
PotentiometricReliability: MediumHandheldLabOn-the-goProxy, needs calibrationSpot reading€€Commercial
One measurement principle for many ions: the same meter reads nitrate, potassium, sodium, calcium or pH depending on which electrode is fitted. Used on soil extracts and soil solution, handheld or in the lab, and in research systems that map pH, K and nitrate together while driving.
MeasuresNO₃⁻KNaCapHP
- How it works
- Potentiometry: a membrane that lets mainly one ion through develops a voltage against a reference electrode, changing by about 59/z mV for every tenfold change in ion activity (Nernst equation). Only the membrane changes from ion to ion. ISFETs put the same membranes on a transistor gate — smaller and faster.
- Range and accuracy
- Validated and sold for pH, NO₃⁻, K⁺, Na⁺ and Ca²⁺. Phosphate electrodes (cobalt-based) responded to H₂PO₄⁻ from 10⁻⁵ to 10⁻¹ M in research but are not field-ready. On the go, integrated pH–K–nitrate mapping worked best for pH, then K, then nitrate.
- Limits
- Other ions interfere (chloride and bicarbonate for nitrate electrodes), readings depend on the soil:water ratio, and each electrode drifts and needs its own calibration. The response becomes non-Nernstian at very low concentrations.