Laboratory soil pH (glass electrode)FR pH du sol au laboratoire (électrode de verre) (pH eau, pH CaCl₂)TR Laboratuvarda toprak pH'ı (cam elektrot) (Toprak pH analizi, su ile pH, CaCl₂ ile pH)

Also called: Soil pH test, pH in water, pH in CaCl₂

PotentiometricReliability: HighLabDirectSpot reading€Reference method

The acidity or alkalinity of a soil sample, measured in a lab suspension. It is the basic soil test: pH controls nutrient availability and is used to decide on liming.

How it works
Dried soil is shaken with water or with a dilute 0.01 M calcium chloride (CaCl₂) solution at a fixed ratio, commonly between 1:1 and 1:5. A glass electrode and a reference electrode develop a voltage that depends on hydrogen-ion activity; the meter converts it to pH.
Range and accuracy
pH units. In CaCl₂ the reading is on average about 0.6 units lower than in water. A multi-laboratory study found reproducibility of roughly 2–11 % relative standard deviation.
Limits
Results depend on the method: water or CaCl₂, the soil-to-solution ratio and the salt content of the sample, so values from different labs are not always comparable. Water pH shifts with seasonal salt levels; the CaCl₂ method is more stable.

Linked fromFertilization › Phosphorus (P), Fertilization › Ammonium sulfate (21-0-0, 24 % S), Fertilization › Soil test interpretation, Fertilization › Soil pH, liming and acidification, Soil › Soil pH and buffering, Soil › Acid soils of the Black Sea coast

Sources: [22] [23] [24]

Sources

  1. Soil Testing: Soil pH and Salt Concentration (Circular 875). University of Georgia Cooperative Extension.
  2. Comparison of Soil pH Methods on Soils of North America. Miller & Kissel, Soil Science Society of America Journal, 2010.
  3. Determination of pH of Soils by Different Methods: Collaborative Study. Y. P. Kalra, Journal of AOAC International, 1995.